maXis UHR-TOF
Bruker
Instrument type:
Qq-TOF
Specifications:
Mass range: m/z 50-20,000
Resolution: up to 40,000
Mass accuracy: < 5 ppm (external) and < 1 ppm (internal)
Ion sources:
Electrospray ionization (ESI)
Nano-electrospray ionization (nano-ESI)
Atmospheric pressure chemical ionization (APCI)
Fragmentation techniques:
Collision-induced dissociation (CID)
Hyphenation:
HPLC
Micro-flow UHPLC
Capillary electrophoreses (CE)
Main applications:
HRMS analysis/QC of small molecules (< 1000 Da)
LC-HRMS analysis for the identification of small molecules in complex samples
Structure elucidation of natural products
Personal remark:
Our working horse for the exact mass determination of synthetic organic compounds, metal-organic/inorganic complexes and natural products.